Review Events Review Page
TM80 Telemetry Monitor Operator's Manual 11 - 5
11.2.4 Changing the Resolution of Trend Data
1. Enter the tabular trends review page.
2. Tap Interval in the lower left corner of the review page.
3. Tap Interval.
4. Select the desired option.
◆ 5 s or 30 s: select to view up to 4 hours of tabular trends at an interval of 5
seconds or 30 seconds.
◆ 1 min, 5 min, 10 min, 15 min, 30 min, 1 h, 2 h, 3 h: select to view up to 48
hours of tabular trends at selected interval.
◆ NIBP: select to view the tabular trends when NIBP measurements are
acquired.
11.3 Events Review Page
The TM80 stores events in real time, including the “MPAN Disconnected” technical alarm
event, the “SpO
2
No Sensor” technical alarm event, physiological alarm events, manual
events, and operational events.
11.3.1 Entering the Events Review Page
Enter the events review page in either of the following ways.
Option 1:
1. Press the key to enter the main menu.
2. Tap Review.
3. Tap Events.
NOTE
• A total loss of power has no impact on the events stored.
• Alarms are saved as events and will be maintained if the equipment is
powered down. The time of equipment power down is not recorded as
an event and cannot be reviewed.
• Earlier events will be overwritten by later ones if the capacity is reached.
• When the TM80 is connected to the CMS, if events of the TM80 are
deleted from the CMS, they will not be deleted from the TM80 and vice
versa.