Monitoring Respiration (Resp) (Optional) Preparation for Monitoring Resp
TM80 Telemetry Monitor Operator’s Manual 8 - 3
8.3 Preparation for Monitoring Resp
8.3.1 Preparing the Patient’s Skin
To prepare the patient’s skin, follow this procedure:
1. Shave the chest hair from the electrode sites in a circular area with a diameter of 2
to 4 inches.
2. Use a dry gauze pad to remove excess skin oils, skin cells and residue from the
electrode sites. Never rub the skin until it is raw or bleeding.
8.3.2 Positioning the Electrodes
As the Respiration measurement adopts the standard ECG electrode placement, you can
use different ECG cables. Since the respiration signal is measured between two ECG elec-
trodes, if a standard ECG electrode placement is applied, the two electrodes should be
RA and LL of ECG Lead II.
For more information, refer to "Standard 3-Leadwire Electrode Placement" on page 7 - 6..
NOTE
• Proper skin preparation is necessary for good signal quality at the elec-
trode site, as the skin is a poor conductor of electricity.
Lead II
RA (R)
LL (F)

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals