Triggering an Alarm Reminding Patients' Location Monitoring with the TM80 at the CMS
13 - 14 TM80 Telemetry Monitor Operator’s Manual
13.10 Triggering an Alarm Reminding Patients' Location
You can enable the functionality of triggering an alarm when patients enter or move out
of the restricted area at the CMS.
After the functionality is enabled, when a patient enter or move out of the restricted
area, the “Out of the allowed area ” alarm will be triggered at the TM80.
To enable the functionality, follow this procedure:
1. Select the system menu area in the upper left corner of the CMS screen.
2. Select System Setup to access the System Setup menu.
3. Select the Network tab.
4. Select the AP Management tab.
5. In the Patient Movement Area Setup section, configure Setup Mode and
Department.
6. If you need to specify an AP location, enter the desired AP location in the Please
input location information field. If you wish to select all the locations, select
Select All.
7. Enable or disable Trigger an alarm when patients enter the restricted area or
Trigger an alarm when patients move out of the allowed area. This option is
dependent on Setup Mode.
8. Select Alarm Priority.
When the “Out of the allowed area ” alarm is triggered, check patient’s location and tap
Confirm at the TM80.