10.5 Interactions after Pairing a TM80 with a BP10
10.5.1 Overview of Interactions
Once a TM80 is paired successfully with a BP10, the interactions proceed as shown
below.
10.5.2 Operating the BP10 via the TM80
When the TM80 is paired successfully with the BP10, you can enter the NIBP menu in
either of the following ways:
â– On the main screen, tap the NIBP digital area.
■Press the key to enter the main menu →tap Parameter Setup → tap NIBP.
In the Actions section of the NIBP menu, you can:
â– Tap Start to start NIBP measurement at the BP10.
â– Tap Stop to stop an ongoing NIBP measurement at the BP10.
â– Tap Stop All to cancel the series measurement in Auto, Sequence, or ABPM mode
at the BP10.
â– Tap NIBP STAT to start the series measurement in continuous mode at the BP10.
Action At the TM80 At the BP10
Start NIBP measurement Yes Yes
Stop NIBP measurement Yes Yes
Stop All NIBP measurements Yes Yes
NIBP STAT Yes Yes
Set initial pressure Yes No
Set NIBP measurement mode Yes No
Configure settings related to Auto
measurement mode
Yes No
Configure settings related to NIBP Sequence
measurement mode
Yes No
Set NIBP settings related to NIBP ABPM
measurement mode
Yes No
Set venipuncture No Yes
Set the cuff pressure for venipuncture No Yes