EasyManua.ls Logo

Mindray BeneVision TM80 - Configuring the ECG Waveform Size; Configuring ECG Alarm Settings

Mindray BeneVision TM80
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Changing ECG Settings Monitoring ECG, Arrhythmia, ST and QT
7 - 16 TM80 Telemetry Monitor Operator’s Manual
2. Tap the icon to exit the ECG menu.
7.4.6 Configuring the ECG Waveform Size
The Waveform Size field of the ECG menu lists all available leads. You can select the
desired ECG lead to set the waveform size. For details about the waveform size setting,
refer to "Configuring ECG Waveforms" on page 7 - 15.
7.4.7 Configuring ECG Alarm Settings
To configure ECG alarm settings, follow this procedure:
1. In the Alarms section of the ECG menu, tap ECG Alarm Setup.The Alarm Limits
menu is displayed.
2. Select the options described in the following table.
Pacer Reject Selects whether or not to reject the pace
pulses.
On, Off
Options Description Settings*
* The factory default settings are in bold.
NOTE
When Paced is set to Yes, the Markers and Pacer Rejection options can
be available.

Table of Contents

Other manuals for Mindray BeneVision TM80

Related product manuals