General Problems Troubleshooting
15 - 2 TM80 Telemetry Monitor Operator’s Manual
15.1 General Problems
The following table lists the problems that are likely to occur. If the problem persists after
corrective actions have been taken, contact your service personnel.
For methods to troubleshoot the central charger, refer to BeneVision Central Charger
Operator’s Manual. For methods to troubleshoot the BP10, refer to BP10 NIBP Module
Operator’s Manual.
Symptom Possible cause Solution
ECG noise The noise interference is overlapped
with ECG waveforms.
1. Check that the electrodes are in good
contact with the skin.
2. Check that the ECG leadwires are
connected securely.
3. Check that the patient does not
contact any ungrounded electric
equipment.
ECG signal saturated The TM80 detected ECG signal satu-
ration or overload.
1. Check the ECG leadwires.
2. Check that the electrodes are in good
contact with the skin.
3. Check that the electrodes are not
expired.
The TM80 or SpO
2
module is started
repeatedly.
The battery capacity for the TM80 is
depleted.
Replace with a known good battery.
The SpO
2
data is not
displayed on the Cen-
tral Station.
1. The SpO
2
module is not
connected to the TM80.
2. There might be a problem
with the SpO
2
module.
1. Connect the SpO
2
module to the
TM80.
2. Replace the SpO2 module with a
known good one.