Unpairing the TM80 with the BP10 Monitoring Noninvasive Blood Pressure (NIBP) (Optional)
10 - 4 TM80 Telemetry Monitor Operator’s Manual
◆ If neither Ye s nor No is selected, the wireless setup confirmation message
will disappear in three minutes. Besides, the “Pairing failure” message will be
displayed in the prompt message area of the TM80. The “MPAN
Disconnected” message is displayed in the message area of the BP10. In this
case, follow steps in Pairing a TM80 with a BP10 on Page 10-2 to pair the
TM80 with the BP10 again.
■ When NIBP measurement mode is set to Manual, Auto, or Sequence and the
BP10 is paired successfully with the TM80, the wireless setup confirmation
message “Select Yes to continue Patient in BP10. Select No to continue Patient in
Telemetry.” will not be displayed. Patient-related settings from the TM80 will be
transferred to the BP10. But NIBP-related settings displayed on the main screen of
the TM80 will come from the BP10. The patient monitored by BP10 will be dis-
charged automatically and the data history related to this patient will be cleared.
■ Patient Category can be changed at the TM80 only.
■ A patient cannot be discharged from the BP10 and NIBP-related settings are
grayed out at BP10. To discharge a patient from BP10, you need to unpair the BP10
form the TM80 by following steps in Unpairing the TM80 with the BP10 on Page
10-4.
10.3 Unpairing the TM80 with the BP10
You can unpair the TM80 with the BP10 either via the TM80 or via the BP10.
10.3.1 Unpairing via the TM80
You can choose either of the two methods to unpair the TM80 with the BP10 via the
TM80.
Option 1:
1. Press the key to enter the main menu of the TM80.
2. Tap Wireless Modules.
3. Select the device you wish to disconnect from the TM80 from the devices list
under Devices.
4. Tap Disconnect. The connected device is disconnected from the TM80.