ST Segment Monitoring Monitoring ECG, Arrhythmia, ST and QT
7 - 32 TM80 Telemetry Monitor Operator’s Manual
Options Description Settings*
ST-I,
ST-II,
ST-III,
ST-aVR,
ST-aVF,
ST-V,
ST-Va,
ST-Vb
These options are displayed
when ST Alarm Mode is set to
Absolute.
The lead label behind ST is con-
sistent with the ECG cable type
selected.
On, Off
Alarm high limit range:
0.2 mV to 2.0 mV; the default is 0.2; the step is
0.01mV.
Alarm low limit range:
-2.0 mV to 0 mV; the default is -0.2; the step is
0.01mV.
Alarm priority: Low, Med, High
ST Single These options are displayed
when ST Alarm Mode is set to
Relative.
On, Off
Alarm high limit range:
0 mV to 2.0 mV; the default is 0.1; the step is
0.01mV.
Alarm low limit range:
-2.0 mV to 0 mV; the default is -0.1; the step is
0.01mV.
Alarm priority: Low, Med, High
ST Dual On, Off
Alarm high limit range:
0 mV to 2.0 mV; the default is 0.1; the step is
0.01mV.
Alarm low limit range:
-2.0 mV to 0 mV; the default is -0.1; the step is
0.01mV.
Alarm priority: Low, Med, High
* The factory default settings are in bold.