Safety Information Safety
1 - 4 TM80 Telemetry Monitor Operator’s Manual
WARNING
Do not open the equipment housings. All servicing and future upgrades
must be carried out by trained and authorized personnel.
Do not rely exclusively on the audible alarm system for patient monitor-
ing. Adjustment of alarm volume to a low level may result in a hazard to
the patient. Always keep the patient under close surveillance.
The physiological data and alarm messages displayed on the TM80
Telemetry Monitor are for reference only and cannot be directly used for
diagnostic interpretation.
PATIENTS WITH A PACEMAKER – On ventricular paced patients, episodes
of ventricular tachycardia may not always be detected. Do not rely
entirely upon the system’s automated arrhythmia detection algorithm.
Keep pacemaker patients under close surveillance.
Do not operate the touch screen with wet hand.
Only use parts and accessories specified in this manual.
Route, wrap and secure the cables to avoid inadvertent disconnection,
stumbling and entanglement.
Instruct patients not to interact with the display of the TM80 Telemetry
Monitor and not to open the battery compartment while the TM80
Telemetry Monitor is in use.
The TM80 Telemetry Monitor should not be used for primary monitoring
in applications where momentary loss of the ECG is unacceptable at the
Central Monitoring System.
We recommend that the WPA2-PSK or WPA2 CCKM security mode be
used when the TM80 Telemetry Monitor is in use.
Auditory alarm signal sound pressure levels that are less than ambient
levels can impede operator recognition of alarm conditions.