EasyManua.ls Logo

R&S ZNL Series - Page 1007

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
1007User Manual 1178.5966.02 ─ 20
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog? <CalKitName>
Queries all calibration standards of a characterization (calkit file) that is stored on a
connected calibration unit, either on the calibration unit's internal (flash) memory or on
an SD card inserted at the calibration unit (if available). The query SYSTem:
COMMunicate:RDEVice:AKAL:CKIT:CATalog? returns the names of the cal kit
files. The factory characterization is always available.
Query parameters:
<CalKitName> Name of the calkit file, string parameter.
'Factory' denotes the factory characterization, the prefix
"SD:" indicates that the characterization is stored on an SD card
inserted at the calibration unit.
Example: See SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
CATalog?
Usage: Query only
Manual operation: See "Characterization" on page 476
SYSTem:COMMunicate:RDEVice:AKAL:DATE? <CalKitName>
Queries the creation date and time of the cal unit characterization (calkit file) <CalKit-
Name>. A possible response is 'Friday, May 26, 2011, 10:13:40'. An empty
string is returned if no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via SYSTem:COMMunicate:RDEVice:AKAL:ADDRess.
Query parameters:
CalKitName String parameter containing the name of a cal unit characteriza-
tion (calkit file).
'Factory' denotes the factory characterization; an empty
string ' ' refers to the last referenced characterization; the pre-
fix "SD:" indicates that the characterization is stored on an SD
card inserted at the calibration unit.
Example: See SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
CATalog?
Usage: Query only
Manual operation: See "Characterization" on page 476
SYSTem:COMMunicate:RDEVice:AKAL:FRANge? <CalKitName>
Queries the frequency range of the cal unit characterization (calkit file) <CalKitName>.
The response consists of the start and stop frequencies in Hz, separated by a comma.
0,0 is returned if no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via SYSTem:COMMunicate:RDEVice:AKAL:ADDRess.
VNA command reference

Table of Contents

Related product manuals