Command reference
R&S
®
ZNL/ZNLE
586User Manual 1178.5966.02 ─ 20
The contents that are reloaded from the file are defined by the last selection made
either in the "Save/Recall" dialogs (manual operation) or through the
MMEMory:SELect[:ITEM] commands (remote operation; the settings are identical in
both cases).
The selection only applies to Spectrum setups, see the R&S ZNL/ZNLE Spectrum
Mode User Manual. A VNA setup is always loaded in its entirety.
By default, the selection is limited to the user settings ("User Settings" selection in the
dialogs, HWSettings in SCPI). The selection is not reset by [Preset] or *RST.
As a consequence, the results of a SCPI script using the MMEMory:LOAD:STATe com-
mand without a previous MMEMory:SELect[:ITEM] command may vary, depending
on previous actions in the GUI or in previous scripts, even if the script starts with the
*RST command.
It is therefore recommended that you use the appropriate MMEMory:SELect[:ITEM]
command before using MMEMory:LOAD:STATe.
Parameters:
<1>
<FileName> String containing the path and name of the file to load.
The string may or may not include the file's extension.
Example:
MMEM:SEL:ALL
//Save all items (User Settings, All Traces, All Limit Lines) from
the R&S ZNL/ZNLE.
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\ZNL\user\TEST01'
//Reloads all items
In the "Recall" dialog, select only "User Settings" and "All Limit
Lines".
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\ZNL\user\TEST01'
//Reloads user settings and all limit lines.
*RST
//Reset instrument.
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\ZNL\user\TEST01'
//Selected items are retained. Reloads user settings and all limit
lines.
Restart the instrument.
(Switch the [ON/OFF] key off and on).
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\ZNL\user\TEST01'
// Selected items are set to default. Reloads only the user set-
tings.
Manual operation:
See "Recall" on page 98
Common instrument commands

Table of Contents

Other manuals for R&S ZNL Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNL Series and is the answer not in the manual?

R&S ZNL Series Specifications

General IconGeneral
BrandR&S
ModelZNL Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals