Command reference
R&S
®
ZNL/ZNLE
587User Manual 1178.5966.02 ─ 20
MMEMory:LOAD:TYPE <Type>
This command defines whether the channel setups that will be loaded with the subse-
quent MMEM:LOAD:STAT command will replace the current channel setup or activate a
new channel setup.
Parameters:
<Type> NEW | REPLace
NEW
The loaded settings will be activated in a new channel setup.
REPLace
The loaded settings will replace the currently active channel
setup.
*RST: NEW
Example:
INST:SEL 'SPECTRUM2'
//Selects channel setup 'SPECTRUM2'.
MMEM:STOR:TYP CHAN
//Specifies that channel setup data is to be stored.
MMEM:STOR:STAT 1, 'C:\Analyzer\Spectrum'
//Stores the settings from channel setup
//'SPECTRUM2' to the file 'C:\Analyzer\Spectrum'.
MMEM:LOAD:TYPE NEW
//Specifies that channel setups are to be loaded
//in a new channel setup.
MMEM:LOAD:STAT 1, 'C:\Analyzer\Spectrum'
//Loads the channel setup from the file
//'C:\Analyzer\Spectrum' to the new channel setup
//'SPECTRUM2*'.
MMEMory:STORe<1|2>:STATe <1>, <FileName>
This command saves the current instrument configuration in a *.dfl file.
Suffix:
<1|2>
.
irrelevant
Parameters:
<1>
<FileName> String containing the path and name of the target file.
The file extension is .dfl.
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current instrument settings in the file Save.dfl.
Manual operation: See "Save File" on page 101
MMEMory:STORe<1|2>:STATe:NEXT
This command saves the current instrument configuration in a *.dfl file.
Common instrument commands