Command reference
R&S
®
ZNL/ZNLE
614User Manual 1178.5966.02 ─ 20
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator <Terminator>
This command selects the GPIB receive terminator.
Output of binary data from the instrument to the control computer does not require
such a terminator change.
Parameters:
<Terminator> LFEOI | EOI
LFEOI
According to the standard, the terminator in ASCII is <LF>
and/or <EOI>.
EOI
For binary data transfers (e.g. trace data) from the control com-
puter to the instrument, the binary code used for <LF> might be
included in the binary data block, and therefore should not be
interpreted as a terminator in this particular case. This can be
avoided by using only the receive terminator EOI.
*RST: LFEOI
Example:
SYST:COMM:GPIB:RTER EOI
Manual operation: See "GPIB Terminator" on page 158
SYSTem:COMMunicate:INTernal:REMote <State>
The instrument switches between manual and remote operation.
Note: If the local lockout function (LLO or SYST:KLOC ON) has been activated in the
remote control mode, manual operation is no longer available until GTL (or SYST:KLOC
OFF) is executed.
For details, see Chapter 8.4.6, "Returning to manual mode ("local")", on page 163.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
The instrument switches from remote to manual operation. You
can operate the instrument locally.
ON | 1
The instrument switches from manual to remote operation.
*RST: 0
Example:
SYST:COMM:INT:REM OFF
The instrument switches from remote to manual operation (cor-
responds to @LOC or selecting the "Local" softkey).
Manual operation: See "Local" on page 163
Common instrument commands