Command reference
R&S
®
ZNL/ZNLE
740User Manual 1178.5966.02 ─ 20
Usage: Event
Options: R&S ZNL-K3
Manual operation: See "Auto Number of Points" on page 370
CALCulate<Chn>:TRANsform:DTFault:SELect <DtfSelectCable>
Selects one of the available (predefined or user-defined) cable types for Distance to
Fault measurements.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<DtfSelectCable> Name of the cable type.
Example:
CALCulate1:TRANsform:DTFault:SELect '5088-HFLR'
Selects cable type "5088-HFLR".
Options: R&S ZNL-K3
Manual operation: See "Cable Type..." on page 371
CALCulate<Chn>:TRANsform:DTFault:STARt <Start>
CALCulate<Chn>:TRANsform:DTFault:STOP <Stop>
Defines the window of the Distance to Fault measurement using its start and stop dis-
tance.
See also CALCulate<Chn>:TRANsform:DTFault:CENTer and
CALCulate<Chn>:TRANsform:DTFault:SPAN.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Start>/<Stop> Start/stop distance.
Default unit: m
Example:
:CALCulate1:TRANsform:DTFault:STARt 1; STOP 3
Defines the DtF distance window of the active trace of channel 1
as [1m, 3m].
Options: R&S ZNL-K3
Manual operation: See "Start Distance / Stop Distance" on page 369
CALCulate<Chn>:TRANsform:DTFault:STATe <DtfState>
If channel <Chn> is performing an unsegmented frequency sweep, this command
allows to transform its active trace into a Distance to Fault (DtF) trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
VNA command reference