Command reference
R&S
®
ZNL/ZNLE
741User Manual 1178.5966.02 ─ 20
Parameters:
<DtfState> OFF|ON – DtF disabled|enabled
*RST: OFF
Example:
:CALCulate1:TRANsform:DTFault:STATE ON
turns the DtF representation of the active trace of channel 1 ON.
Options: R&S ZNL-K3
Manual operation: See "Distance to Fault" on page 368
CALCulate:TRANsform:TIME...
The CALCulate:TRANsform:TIME... commands control the transformation into the
time domain (see Chapter 9.7.1, "Time domain analysis", on page 294).
CALCulate<Chn>:TRANsform:TIME:CENTer.................................................................... 741
CALCulate<Chn>:TRANsform:TIME:DCHebyshev............................................................ 742
CALCulate<Chn>:TRANsform:TIME:LPASs......................................................................742
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam.....................................................743
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:CONTinuous.................................743
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:EXTRapolate................................ 744
CALCulate<Chn>:TRANsform:TIME:LPFRequency...........................................................744
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor.................................................. 744
CALCulate<Chn>:TRANsform:TIME:SPAN.......................................................................745
CALCulate<Chn>:TRANsform:TIME:STARt...................................................................... 745
CALCulate<Chn>:TRANsform:TIME:STATe...................................................................... 745
CALCulate<Chn>:TRANsform:TIME:STIMulus..................................................................746
CALCulate<Chn>:TRANsform:TIME:STOP.......................................................................746
CALCulate<Chn>:TRANsform:TIME[:TYPE]..................................................................... 747
CALCulate<Chn>:TRANsform:TIME:WINDow...................................................................747
CALCulate<Chn>:TRANsform:TIME:XAXis.......................................................................748
CALCulate<Chn>:TRANsform:TIME:CENTer <CenterTime>
Defines the center time of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<CenterTime> Center time of the diagram in time domain.
Range: -99.99999999999 s to +99.99999999999 s
Increment: 0.1 ns
*RST: 1.5E-009 s
Default unit: s
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:CENT 0; SPAN 5ns
Set the center time to 0 ns and the time span to 5 ns.
VNA command reference