EasyManua.ls Logo

R&S ZNL Series - Page 741

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
741User Manual 1178.5966.02 ─ 20
Parameters:
<DtfState> OFF|ON – DtF disabled|enabled
*RST: OFF
Example:
:CALCulate1:TRANsform:DTFault:STATE ON
turns the DtF representation of the active trace of channel 1 ON.
Options: R&S ZNL-K3
Manual operation: See "Distance to Fault" on page 368
CALCulate:TRANsform:TIME...
The CALCulate:TRANsform:TIME... commands control the transformation into the
time domain (see Chapter 9.7.1, "Time domain analysis", on page 294).
CALCulate<Chn>:TRANsform:TIME:CENTer.................................................................... 741
CALCulate<Chn>:TRANsform:TIME:DCHebyshev............................................................ 742
CALCulate<Chn>:TRANsform:TIME:LPASs......................................................................742
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam.....................................................743
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:CONTinuous.................................743
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:EXTRapolate................................ 744
CALCulate<Chn>:TRANsform:TIME:LPFRequency...........................................................744
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor.................................................. 744
CALCulate<Chn>:TRANsform:TIME:SPAN.......................................................................745
CALCulate<Chn>:TRANsform:TIME:STARt...................................................................... 745
CALCulate<Chn>:TRANsform:TIME:STATe...................................................................... 745
CALCulate<Chn>:TRANsform:TIME:STIMulus..................................................................746
CALCulate<Chn>:TRANsform:TIME:STOP.......................................................................746
CALCulate<Chn>:TRANsform:TIME[:TYPE]..................................................................... 747
CALCulate<Chn>:TRANsform:TIME:WINDow...................................................................747
CALCulate<Chn>:TRANsform:TIME:XAXis.......................................................................748
CALCulate<Chn>:TRANsform:TIME:CENTer <CenterTime>
Defines the center time of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<CenterTime> Center time of the diagram in time domain.
Range: -99.99999999999 s to +99.99999999999 s
Increment: 0.1 ns
*RST: 1.5E-009 s
Default unit: s
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:CENT 0; SPAN 5ns
Set the center time to 0 ns and the time span to 5 ns.