Command reference
R&S
®
ZNL/ZNLE
779User Manual 1178.5966.02 ─ 20
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:R<1|2|3> <CircuitModel>[, <Resistance>]
Specifies the resistance value R<i> in the different lumped circuit models for port pair
deembedding.
In the query form, the <Resistance> parameter must be omitted. The command returns
the resistance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port pair (see CALCulate<Ch>:
TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
DEFine)
<1|2|3> Index i of the resistance R<i> in the related lumped circuit
model. If unspecified the numeric suffix is set to 1.
Parameters:
<CircuitModel> STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL |
SLCS | SCLS | SCCS | SLLS | GSSL | LSSG | SLGS | SGLS
Circuit model whose resistance R<i> shall be set, see
Table 11-10
<Resistance> Resistance R<i> for the specified circuit model.
Range: -10 MΩ to 10 MΩ
Increment: 1 mΩ
*RST: 0 Ω if the resistance is connected in series with an
inductance. 10 MΩ if the resistance is connected in
parallel with a capacitance)
Default unit: Ohm
Example: See CALCulate<Ch>:TRANsform:VNETworks:PPAir:
DEEMbedding<ListId>[:STATe]
Manual operation: See "Network" on page 553
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>[:STATe]
<arg0>
Enables or disables the deembedding function for port set (or port pair) <ListId>. It is
allowed to change the deembedding network while embedding is enabled.
Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port set
Parameters:
<arg0> OFF (0): Deembedding inactive
ON (1): Deembedding active
*RST: OFF (0)
VNA command reference

Table of Contents

Related product manuals