Command reference
R&S
®
ZNL/ZNLE
888User Manual 1178.5966.02 ─ 20
Manual operation: See "Add / Copy / Delete / Standards..." on page 500
[SENSe:]CORRection:CKIT:DMODe <ConnectorType>, <CalKitName>, <KitLabel>,
<Mode>
Sets/gets the delay mode for the related cal kit (identified by connector type, name and
label), i.e allows to toggle between ZVR compatible and Keysight modelling (see "Off-
set Parameters" on page 503). Subsequent standard definitions interpret the specified
<DelayParam> accordingly.
In "set direction", if a cal kit with the given connector type, name and label is not availa-
ble on the analyzer, it is created automatically.
Cal kits that are not created/modified using [SENSe:]CORRection:CKIT:DMODe use
the ZVR compatible modelling.
Parameters:
<ConnectorType> Name of the connector type. Use [SENSe<Ch>:
]CORRection:CONNection:CATalog? to query connector
names.
<CalKitName> String parameter containing the name of the calibration kit. See
"Cal kit naming conventions " under [SENSe:]CORRection:
CKIT:<ConnType>:LSELect.
<KitLabel> String parameter containing the label of the calibration kit, usu-
ally the serial number.
Setting parameters:
<Mode> DELay | ELENgth
DELay – Keysight modelling
ELENgth – ZVR compatible modelling
Manual operation: See "Add / Copy / Delete / Standards..." on page 500
[SENSe:]CORRection:CKIT:LABel <CalKitName>[, <KitLabel>]
Assigns a label to an imported or user-defined calibration kit.
Parameters:
<CalKitName> String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
<KitLabel> String parameter containing the calibration kit label.
VNA command reference