Command reference
R&S
®
ZNL/ZNLE
889User Manual 1178.5966.02 ─ 20
Example:
CORR:CKIT:FOP 'N 50 Ohm','New Kit 1', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Create a new cal kit "New Kit 1" and assign an open (f) standard
for the N 50 Ω connector type with specific properties.
CORR:CKIT:LAB 'New Kit 1', 'Test kit created
today'
Label the previously created kit.
CORR:CKIT:LAB? 'New Kit 1'
Check the label.
CORR:CKIT:DEL 'New Kit 1'
Delete the kit.
[SENSe:]CORRection:CKIT:SELect <ConnectorType>[, <CalKitName>]
Selects the calibration kit to be used, specifying its connector type and name
(optional).
Tip: The command is suitable for connector types with arbitrary, user-defined names.
For standard connector types you can use the command [SENSe:]CORRection:
CKIT:<ConnType>:SELect.
Parameters:
<ConnectorType> Connector type, e.g. a user-defined connector type (string vari-
able).
<CalKitName> String parameter containing the name of a calibration kit availa-
ble on the analyzer.
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\
Documents\Rohde-Schwarz\ZNL\
Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file New Kit 1.calkit
from the default cal kit directory.
CORR:CKIT:SEL 'N 50 Ohm', 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in New Kit 1.calkit reads New
Kit 1.
Manual operation: See "Cal Kit" on page 484
[SENSe:]CORRection:CKIT:<ConnType>:SELect <CalKitName>
Selects the calibration kit to be used for a specified connector type <ConnType>. The
kit is identified by its name.
Tip: For connector types with arbitrary, user-defined names you can use the command
[SENSe:]CORRection:CKIT:SELect.
Cal kit naming conventions
VNA command reference