Command reference
R&S
®
ZNL/ZNLE
902User Manual 1178.5966.02 ─ 20
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs <Characterization>,
<TestPort1>, <CalUnitPort1>[, <TestPort2>...]
Generates a characterization (cal kit file) with the specified name containing the cal kit
data of the active calibration unit (SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
). The cal kit file can be saved to a directory on the analyzer, to the calibration unit's
internal (flash) memory or to an SD card inserted at the calibration unit (if available).
Similar logic as [SENSe:]CORRection:COLLect:AUTO:CKIT, but with flexible port
assignment.
Setting parameters:
<Characterization> Location of the created characterization / cal kit file (string
parameter, extension *.calkit):
If a path is specified, the file is saved to the analyzer's hard disk.
The default directory (MMEMory:CDIRectory) is not used.
If only the file name is specified, the file is saved to the calibra-
tion unit's internal (flash) memory. The factory calibration data on
the unit is not overwritten.
If the file name is prefixed by "SD:", the file is saved to the SD
card inserted at the calibration unit (if available/accessible).
<TestPort1> Number of first test port.
<CalUnitPort1> Number of the calibration unit port to whom the first test port
(<TestPort1>) is assigned.
<TestPort2> Number of second test port...
Usage: Setting only
Manual operation: See "Test Port Assignment" on page 507
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs:ADD <Characterization>,
<TestPort1>, <CalUnitPort1>[, <TestPort2>]...
Extends or modifies an existing characterization of the active calibration unit (SYSTem:
COMMunicate:RDEVice:AKAL:ADDRess).
This functionality is not available at the GUI.
Setting parameters:
<Characterization> Location of an existing characterization (e.g. created using
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs).
A VNA cal kit file name *.calkit without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
A VNA cal kit file name *.calkit with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1> Number of first test port.
VNA command reference

Table of Contents

Other manuals for R&S ZNL Series