EasyManua.ls Logo

R&S ZNL Series - Page 980

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
980User Manual 1178.5966.02 ─ 20
Example: See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 983
Usage: Query only
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 515
SOURce:POWer:CORRection:TCOefficient:DEFine<ListNo> <Frequency>,
<TransCoeff>[, <DeEmbedding Position>]
Adds a new frequency and transmission coefficient to the end of the transmission coef-
ficient list. The query returns the frequency and transmission coefficient no. <ListNo>.
Suffix:
<ListNo>
.
List entry no. (only used for queries)
Parameters:
<Frequency> Stimulus frequency value. If several points with identical fre-
quencies are added, the analyzer automatically ensures a fre-
quency spacing of 1 Hz.
Range: Stimulus values outside the frequency range of the
analyzer are allowed.
*RST: n/a
Default unit: Hz
<TransCoeff> Transmission coefficient
Range: -300 dB to +200 dB
*RST: n/a
Default unit: dB
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Example:
See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 983
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 515
VNA command reference

Table of Contents

Related product manuals