R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 987
Device-specific, command or query
FREQ:STAR 0.9E+9; STOP 1.1E+9
Set frequency range from 0.9 GHz to 1.1 GHz.
SWEep:POINts 101
Set number of sweep points to 101 resulting in a frequency spacing of 2
MHz.
SOUR:GRO 1,2
SOUR:GRO 3,4
Create two port groups with ports 1 and 2 in port group 1 and ports 3 and 4 in
port group 2.
SOURce:GROup:SIMultaneous:FOFFset:MOFFset:BWFactor 15
Set minimum frequency offset to 15 times the maximum bandwidth in the
sweep.
SOURce:GROup:SIMultaneous:FOFFset:STATe ON
Activate Simultaneous Measurement with Frequency Offset.
SOURce:GROup:SIMultaneous:FOFFset:CONDition?
Check the operation mode of Simultaneous Measurement with Frequency
Offset.
SOURce<Ch>:GROup:SIMultaneous:FOFFset:MOFFset:DVALue <min_offset>
This command is used for Simultaneous Measurement with Frequency Offset to directly specify the
minimum frequency offset in DIRect mode.
If SOURce<Ch>:GROup:SIMultaneous:FOFFset:MOFFset:MODE is set to BANDwidth, the value set
via this command has no effect. For details see description of remote command
SOURce<Ch>:GROup:SIMultaneous:FOFFset:MOFFset:MODE.
Channel number.
If unspecified, this parameter is set to 1.
The minimum frequency offset
0 Hz to 70 MHz [Hz]
A value of 0 Hz will result in an actual offset equal to the frequency spacing
between consecutive measurement points.
Device-specific, command or query
See SOURce<Ch>:GROup:SIMultaneous:FOFFset[:STATe]
SOURce<Ch>:GROup:SIMultaneous:FOFFset:MOFFset:MODE DIRect |