Replace the previous 3 points by the trace points of the default trace 'Trc1'.
MMEM:STOR:CORR:TCO 'C:\Program Files\Rohde&Schwarz\Network
Analyzer\PowermeterCorrLists\Test.pmcl'
Store the power loss list to the power meter correction list file Test.pmcl in the
directory C:\Program Files\Rohde&Schwarz\Network
Analyzer\PowermeterCorrLists.
MMEM:LOAD:CORR:TCO 'C:\Program Files\Rohde&Schwarz\Network
Analyzer\PowermeterCorrLists\Test.pmcl'
Re-load the stored power meter correction list file.
SOUR:POW:CORR:TCO:DEL1
Delete the first point in the list.
SOUR:POW:CORR:TCO:DEL:ALL; :SOUR:POW:CORR:TCO?
Clear the entire list. Query whether the transmission coefficients are still taken into
account. The response is 0.
MMEM:LOAD:CORR:TCO 'C:\Program Files\Rohde&Schwarz\Network Analyzer\Traces\Test.s2p', 'S21'
Load a power loss list from a previously created 2-port Touchstone file.
SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:COUNt?
Queries the number of frequency values and transmission coefficients in the power loss list.
Calibrated channel number. This parameter is ignored; the transmission
coefficient settings are valid for all channels.
Calibrated port number. This parameter is ignored; the transmission coefficient
settings are valid for all sources.
Device-specific, query only
See SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration
SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:DEFine<List_No>
<frequency>, <transmission>
Adds a new frequency and transmission coefficient to the end of the power loss list.
Calibrated channel number. This parameter is ignored; the transmission
coefficient settings are valid for all channels.
Calibrated port number. This parameter is ignored; the transmission coefficient
settings are valid for all sources.
Number of point in the list. This suffix is only used for queries.
Stimulus values outside the frequency range of the analyzer are allowed [Hz]. If
several points with identical frequencies are added, the analyzer automatically