STAT:QUES:LIM1?
Query the EVENt part of the QUEStionable:LIMit1 register to check whether an
event has occurred since the last reading.
STATus:QUEStionable:LIMit<1|2>:NTRansition
Sets the negative transition filter. If a bit is set, a 1 to 0 transition in the corresponding bit of the associated
condition register causes a 1 to be written in the associated bit of the corresponding event register.
0 to 65535 (decimal representation)
– (see also Reset Values of the Status Reporting System)
Confirmed, command or query
STAT:QUES:LIM2:NTR 6
Set bits no. 1 and 2 of the QUEStionable:LIMit2:NTRansition register
STATus:QUEStionable:LIMit<1|2>:PTRansition
Sets the positive transition filter. If a bit is set, a 0 to 1 transition in the corresponding bit of the associated
condition register causes a 1 to be written in the associated bit of the corresponding event register.
0 to 65535 (decimal representation)
– (see also Reset Values of the Status Reporting System)
Confirmed, command or query
STAT:QUES:LIM2:PTR 6
Set bits no. 1 and 2 of the QUEStionable:LIMit2:PTRansition register
STATus:QUEue[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is identical to that of
SYSTem:ERRor[:NEXT]?
The entry consists of an error number and a short description of the error. Positive error numbers are
instrument-dependent. Negative error numbers are reserved by the SCPI standard; see section Error
Messages.
STAT:QUE?
Query the oldest entry in the error queue. 0,"No error" is returned if the error