R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1072
functionality has been assigned yet.
SYST:USER:KEY:FUNC 1, 'S11'
Assign the functionality 'S11' (select S11 as a measured quantity for the active
trace) to the created user key.
SYST:USER:KEY?
The query returns 0,'', indicating that no user key has been pressed. If you
press the user softkey no. 1, the response is 1, 'User S11'.
CALC:PAR:MEAS? 'Trc1'
The query returns 'S11','', indicating that the measured quantity for trace 'Trc1'
has been changed.
SYST:USER:KEY 0
Delete the user key and restore the default keys..
SYSTem:USER:KEY:FUNCtion <ukey_no>[, '<ukey_name>']
Labels the functionality of a function key to a user-defined key in the remote display.
You can use this command to execute part of your measurement task manually; see Combining
Manual and Remote Control.
1 to 8 – user key numbers
– (the parameter is also needed in the query form)
Name of a function key, e.g. 'S11', 'Start' etc. (string variable)
Device-specific, command or query.
SYSTem:VERSion?
Returns the SCPI version number to which the analyzer complies. The analyzer complies to the final SCPI
version 1999.0
SYST:VERS?
Query the SCPI version. The response is 1999.0.