R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1081
TRIGger<Ch>[:SEQuence]:TIMer <numeric_value>
Sets the period of the internal periodic signal that can be used as a trigger source
(TRIGger<Ch>[:SEQuence]:SOURce TIMer).
1E-005 s to 13680 s [s]. UP and DOWN increment/decrement the timer
period in 10 ms steps.
Confirmed, command or query
TRIG:SOUR TIM
Activate internal timer as trigger source.
TRIG:TIM 0.1
Set a timer period of 100 ms.