R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1080
Trigger pulsed measurement by rising edge of external pulse signal
TRIGger<Ch>[:SEQuence]:SLOPe POSitive | NEGative
Qualifies whether the trigger event occurs on the rising or on the falling edge of the external TTL trigger
signal.
Confirmed, command or query
TRIG:SOUR EXT
Activate external signal as trigger source.
TRIG:SLOP NEG
Trigger on the negative edge of the (external TTL) trigger signal.
TRIGger<Ch>[:SEQuence]:SOURce IMMediate | EXTernal | TIMer | MANual |
PGENerator
Selects the source for the events that the analyzer uses to start a measurement sequence.
Free run measurement (untriggered)
Trigger by external signal applied to the EXT TRIGGER or pin 2 of USER
CONTROL connector on the rear panel.
Periodic internal trigger signal. The period of the timer is specified by the
TRIGger[:SEQuence]:TIMer command.
Trigger event generated by pressing the Manual Trigger softkey.
Trigger event generated the pulse generator (with option R&S ZVA-K27),
depending on the TRIGger<Ch>[:SEQuence]:PULSe settings.
Confirmed, command or query.
TRIG:SOUR MAN
Activate manual trigger mode. The analyzer starts the next sweep when the
Manual Trigger softkey is pressed.