R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Programming Examples
Condensed Programming Examples
Operating Manual 1145.1084.12 – 30 1122
// Set format for data transfer
:FORMAT REAL,32
//
// !!! Important !!! Stop sweep when loading error terms
:INITiate:CONTinuous OFF
//
// Recall error terms
// CALCulate:DATA is suitable for 2-port terms;
// use SENSe:CORRection:CDATa for more than 2 ports
:CALCulate1:DATA SCORR1,#@scorr1.dat
:CALCulate1:DATA SCORR2,#@scorr2.dat
:CALCulate1:DATA SCORR3,#@scorr3.dat
// We are omitting the isolation term, as it is not implemented on the ZVAB
//:CALCulate1:DATA SCORR4,#@scorr4.dat
:CALCulate1:DATA SCORR5,#@scorr5.dat
:CALCulate1:DATA SCORR6,#@scorr6.dat
:CALCulate1:DATA SCORR7,#@scorr7.dat
:CALCulate1:DATA SCORR8,#@scorr8.dat
:CALCulate1:DATA SCORR9,#@scorr9.dat
// We are omitting the isolation term, as it is not implemented on the ZVAB
//:CALCulate1:DATA SCORR10,#@scorr10.dat
:CALCulate1:DATA SCORR11,#@scorr11.dat
:CALCulate1:DATA SCORR12,#@scorr12.dat
:INITiate:CONTinuous ON
Using a Calibration Data Recorded Previously
You can re-use any set of correction data that you acquired in earlier sessions on your analyzer. In
general you have to carry out the following steps:
1. Create a dummy correction data set and store it to a file.
2. Replace the dummy data with your correction data.
3. Re-import the correction data file and apply it to a channel.
See also the program examples for the following commands:
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault (dummy system error
correction)
SOURce<Ch>:POWer<Pt>:CORRection:DEFault (dummy source power calibration without
external power meter)