shift between the analyzer and the calibration plane).
CORR:DATA 'SCORR3',<ASCII_data>
Replace the dummy system error correction term with your own correction data,
transferred in ASCII format.
INIT:CONT ON
Restart the sweep in continuous mode.
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected
Calculates the system error correction data from the acquired measurement results
([SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected) , stores them, and applies them to
the calibrated channel <Ch>.
For successful execution of this command, complete measurement data sets must be available for all
currently selected calibrations. See the description of
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine) for background information.
To avoid incompatibilities, older system error correction data is deleted if one of the following applies:
The older calibration has the same name as the current calibration (see
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine)
The older calibration is completely covered by the current calibration.
To avoid inadvertent loss of calibration data, you can transfer the active channel calibration to the cal pool
(MMEMory:STORe:CORRection <Ch>, '<file_name>').
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as calibration
type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the measurement
results of this standard.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the active channel.
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault
Generates a set of default system error correction data for the selected ports and calibration type. The
default data set corresponds to a test setup which does not introduce any systematic errors; none of the
measurement results acquired previously
([SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected) is taken into account.