R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 791
Name of a user-defined calibration kit available on the analyzer.
It is not possible to modify or store predefined or ideal kits.
Label of the calibration kit, usually its serial number.
String parameter to specify the name and directory of the cal kit file to be created.
The file is a NWA-specific cal kit file with the extension *.calkit. If no path is
specified the analyzer uses the current directory, to be queried with
MMEMory:CDIRectory?.
Device-specific, no query.
See [SENSe<Ch>:]CORRection:CKIT:LLABel
MMEMory:STORe:CMAP '<file_name>'
Stores a color scheme to a specified NWA color scheme file.
String parameter to specify the name and directory of the color scheme file to be
created. If no path is specified the analyzer uses the current directory, to be
queried with MMEMory:CDIRectory?. The default extension (manual control) for
color scheme files is *.ColorScheme, although other extensions are allowed.
Device-specific, no query.
MMEMory:STORe:CORRection <Ch>,'<file_name>'
Copies the correction data of channel <Ch> to the cal pool, generating a new correction data file (cal
group). The file has the extension *.cal and is stored in the C:\Program
Files\Rohde&Schwarz\Network Analyzer\Calibration\Data directory.
String parameter to specify the name of the created cal group file. There is no
need to specify the directory path and file extension; the analyzer uses the default
cal pool directory C:\Program Files\Rohde&Schwarz\Network
Analyzer\Calibration\Data and a *.cal extension.
Device-specific, no query.
See MMEMory:LOAD:CORRection