R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 828
String parameter containing the label of an imported or user-defined
calibration kit available on the analyzer.
– (A *RST does not affect calibration kits)
Device-specific, no query
See [SENSe<Ch>:]CORRection:CKIT:LLABel
[SENSe<Ch>:]CORRection:CKIT:LLABel '<ckit_name>', '<ckit_label>'[,
'<new_label>'] [SENSe<Ch>:]CORRection:CKIT:LLABel? '<ckit_name>',
'<ckit_label>'
Assigns a calibration kit label to an imported or user-defined calibration kit or renames an existing
calibration kit label.
Channel number. This suffix is ignored because calibration kits are channel-
independent.
String parameter containing an imported or user-defined calibration kit available
on the analyzer.
String parameter containing the current label.
String parameter containing the new calibration kit label.
– (A *RST does not affect calibration kits)
Device-specific, command or query
CORR:CKIT:FOP:WLAB 'N 50 Ohm','New Kit 1','Test kit created today', '',0,4000000000,0,0,50,99,-
2.3,0.22,0,0,0,0,0
Create a new cal kit New Kit 1 labeled "Test kit created today" and assign an
open (f) standard for the N 50 Ω connector type with specific properties..
CORR:CKIT:LLAB 'New Kit 1', 'Test kit created today', '2012-05-25'
Change the label of the previously created kit.
CORR:CKIT:LLAB? 'New Kit 1', '2012-05-25'
Check the label.
MMEMory:STORe:CKIT:WLABel 'New Kit 1', '2012-05-25', 'C:\Program
Files\Rohde&Schwarz\Network Analyzer\Calibration\Kits\New Kit 1 (2012-05-
25).calkit'
Store the data for the labeled cal kit to the cal kit file New Kit 1 (2012-05-
25).calkit.
CORR:CKIT:LDEL 'New Kit 1', '2012-05-25'
Delete the kit from the internal memory.
MMEMory:LOAD:CKIT 'C:\Program Files\Rohde&Schwarz\Network
Analyzer\Calibration\Kits\New Kit 1 (2012-05-25).calkit'
Re-load the kit.
[SENSe<Ch>:]CORRection:CKIT:LSELect '<conn_type>', '<ckit_name>',