Operating Manual 1145.1084.12 – 30 827
N50, N75: N 50 Ω or N 75 Ω connectors
PC7, PC35, PC292: PC 7, PC 3.5 or 2.92 mm connectors
USER<no>: User-defined connectors UserConn1, UserConn2
SMA: User-defined connector type SMA ...
String parameter containing the name of a calibration kit available on the
analyzer.
String parameter containing the label of a calibration kit available on the analyzer,
usually the serial number.
– (A *RST does not change the assignment between connector types and
calibration kits.)
Device-specific, command or query
MMEM:LOAD:CKIT 'C:\Program Files\Rohde&Schwarz\Network
Analyzer\Calibration\Kits\New_kit (123456).calkit'
Load the previously created cal kit file New_kit
(123456).calkit from the default cal kit directory.
CORR:CKIT:N50:LSEL 'New_kit', '123456'
Assign the imported kit to the N 50 Ω connector type (assuming that the cal kit
name stored in New_kit (123456).calkit reads New_kit and that its label
reads 123456).
[SENSe<Ch>:]CORRection:CKIT:LCATalog? '<conn_name>'
Returns a list of all cal kits and their labels for a given connector type.
Channel number. This suffix is ignored because cal kits are channel-
independent.
Name of the connector type. Use
SENSe<Ch>:CORRection:CONNection:CATalog? to query connector
names.
Device-specific, query only
See SENSe<Ch>:CORRection:CONNection:CATalog?
[SENSe<Ch>:]CORRection:CKIT:LDELete '<ckit_name>', '<ckit_label>'
Deletes an imported or user-defined cal kit which is identified by its cal kit name and label.
It is not possible to modify or store predefined or ideal kits.
Channel number. This suffix is ignored because calibration kits are channel-
independent.
String parameter containing an imported or user-defined calibration kit
available on the analyzer.