R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 792
MMEMory:STORe:CORRection:TCOefficient '<file name>'
Saves the current power loss list to a specified power meter correction list file.
String parameter to specify the name and directory of the created power meter
correction list file. The file extension *.pmcl is mandatory. If no path is specified the
analyzer uses the current directory, to be queried with MMEMory:CDIRectory?.
See SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration
MMEMory:STORe:LIMit '<trc_name>', '<file_name>'
Saves the limit lines associated to a specified trace to a limit line file. Limit lines are created using the
CALCulate<Chn>:LIMit... commands.
Name of an existing trace in the active setup (string parameter) for which a limit
line definition exists.
String parameter to specify the name and directory of the created limit line file. The
default extension (manual control) for limit line files is *.limit, although other
extensions are allowed. If no path is specified the analyzer uses the current
directory, to be queried with MMEMory:CDIRectory?.
Device-specific, no query.
MMEMory:STORe:MARKer '<file_name>'
Saves the values of all markers to a ASCII file.
String parameter to specify the name and directory of the created ASCII file. The
default extension (manual control) for marker files is *.txt, although other
extensions are allowed. If no path is specified the analyzer uses the current
directory, to be queried with MMEMory:CDIRectory?. An example for the file
contents is given below.
Device-specific, no query.