See MMEMory:SETTings:RENorm:MODE
MMEMory:SETTings:RENorm:STATe <Boolean>
Enables or disables renormalization during Touchstone file export and import.
This setting is valid in explicit mode only (MMEMory:SETTings:RENorm:MODE EXPLicit).
ON | OFF – enable or disable renormalization.
Device-specific, command or query
See MMEMory:SETTings:RENorm:MODE
MMEMory:STORe:CKIT '<ckit_name>', '<file_name>'
Stores the data of a calibration kit to a specified file. The calibration kit is identified by its name.
Name of a user-defined calibration kit available on the analyzer.
It is not possible to modify or store predefined or ideal kits.
String parameter to specify the name and directory of the cal kit file to be created.
The file is a NWA-specific cal kit file with the extension *.calkit. If no path is
specified the analyzer uses the current directory, to be queried with
MMEMory:CDIRectory?.
Device-specific, no query.
MMEM:LOAD:CKIT 'C:\Program Files\Rohde&Schwarz\Network
Analyzer\Calibration\Kits\New_kit.calkit'
Load the previously created cal kit file New_kit.calkit from
the default cal kit directory.
... :MMEM:STOR:CKIT 'New_kit', 'C:\Program
Files\Rohde&Schwarz\Network
Analyzer\Calibration\Kits\New_kit.calkit'
Store the data for the user-defined cal kit New_kit and
overwrite the cal kit file New_kit.calkit.
MMEMory:STORe:CKIT:WLABel '<ckit_name>', '<ckit_label>', '<file_name>'
Stores the data of a calibration kit to a specified file. The calibration kit is identified by its name and label.