Device-specific, command or query
Suppose that the active setup contains an active trace no. 1.
CALC:MARK:REF ON; :CALC:MARK ON
Create the reference marker and marker 1 and assign them to trace no. 1. The
default position of both markers is the center of the sweep range.
CALCulate<Chn>:MARKer<Mk>:REFerence:TYPE NORMal | FIXed
Sets the reference to normal or fixed mode. The marker must be created before using
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON.
Channel number used to identify the active trace
Marker number in the range 1 to 10. This numeric suffix is ignored and may
be set to any value.
NORMal – response value changes according to the measurement result.
FIXed – marker keeps its current response value.
Device-specific, command or query
CALC:MARK:REF ON; :CALC:MARK:REF:TYPE FIX
Create the reference markerand display it in the center of the sweep range
as a fixed marker.
CALC:MARK:REF:X 1GHz
Shift the marker horizontally. The response value remains fixed.
CALCulate<Chn>:MARKer<Mk>:REFerence:X <stimulus>
Defines the stimulus (in Cartesian diagrams: x-axis) value of the reference marker, which can (but doesn't
have to) be displayed using CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON.
Channel number used to identify the active trace
Marker number in the range 1 to 10. This numeric suffix is ignored and may be
set to any value.
Stimulus value of the reference marker
–9.9E+11 Hz to +9.9E+11 Hz [Hz] (for frequency sweeps)
-999 dBm to +999 dBm [dBm] (for power sweeps)
0 s to 127500 s [s] (for time and CW mode sweeps)
Device-specific, command or query
Suppose that the active setup contains an active trace no. 1 and that the sweep