is FRTR0104.
CORR:COLL:SAVE:SEL:DEF; :MMEM:STOR:CORR 1, 'P1-P4.cal'
Create a default calibration data set for the selected
calibration and store the data to a cal group file.
MMEM:LOAD:CORR:MERGE 1, 'P1-P3.cal', 'P1-P4.cal';
:CORR:COLL:METH:DEF?
Load and merge the two cal group files and query the active
calibrations again. The response is FRTR0103,FRTR0104.
MMEM:STOR:CORR 1, 'Merged.cal'
Store both calibrations to a common cal group file.
MMEMory:LOAD:CORRection:RESolve <channel> | ALL [,'<file_name>']
Resolves the pool link between channel <Ch> and a correction data set (cal group file). After resolving the
pool link, the analyzer keeps the previous system error correction as a channel calibration (Channel Cal).
A new calibration will replace the channel calibration but not overwrite the old cal group file (and not affect
other channels).
Channel number of an existing channel. ALL resolves the pool link for all channels.
Optional string parameter to specify the name of the cal group file. Cal group files
must have the extension *.cal. The directory must not be specified; the analyzer
always uses the default path C:\Program Files\Rohde&Schwarz\Network
Analyzer\Calibration\Data.
If there is no link between <Ch> and the specified file, the command has no effect.
If no file is specified, the command resolves any link between <Ch> and an
arbitrary cal group file.
Device-specific, no query.
See MMEMory:LOAD:CORRection
MMEMory:LOAD:CORRection:TCOefficient '<file name>'[,'<parameter name/trace
name>']
Loads trace data from a specified power meter correction file or trace file and assigns it to a trace with a
specified name. Traces are created using the CALCulate<Ch>:PARameter:SDEFine... command.
Renormalization rules for imported Touchstone files
In automatic renormalization mode (MMEMory:SETTings:RENorm:MODE AUTO), renormalization is
enabled. In explicit renormalization mode (MMEMory:SETTings:RENorm:MODE EXPLicit),
renormalization is controlled by MMEMory:SETTings:RENorm:STATe.
The analyzer uses the connector impedance of the physical ports as a target impedance; the