R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 940
Select pulse profile mode and select an IF bandwidth of 5 MHz.
PULS:TIME:STAR -10 us; STOP 200 us
Select a displayed time range between –10 μs and +200 μs.
[SENSe<Ch>:]PULSe:TIME:STARt <start>
Selects the start time of the displayed time range relative to the trigger time.
–0.003 s to +0.002999975 s [s]. The increment (UP/DOWN) is 50 ns. The width of
the displayed time range must be between 25 ns and 3 s.
Device-specific, command or query
See [SENSe<Ch>:]PULSe:TIME:BWIDth[:RESolution]
[SENSe<Ch>:]PULSe:TIME:STOP <stop>
Selects the stop time of the displayed time range relative to the trigger time.
–0.002999975 s to +0.003 s [s]. The increment (UP/DOWN) is 50 ns. The width of
the displayed time range must be between 25 ns and 3 s.
Device-specific, command or query
See [SENSe<Ch>:]PULSe:TIME:BWIDth[:RESolution]
[SENSe<Ch>:]ROSCillator...
This subsystem controls the frequency reference signal.
[SENSe<Ch>:]ROSCillator[:SOURce] INTernal | EXTernal
Selects the source of the reference oscillator signal.
Channel number. This suffix is ignored in the ROSCillator subsystem and can be