R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 939
Number of the measurement receiver (= port number of the analyzer). Values
range from 1 to the total number of test ports.
Number of the analyzer source port. 1 to the total number of test ports.
ON | OFF – Display or hide lines
Device-specific, command or query
See [SENSe<Ch>:]PULSe:RECeiver:B<rec_no>:SRCPort<gen_no>:EVALuation:MODE
[SENSe<Ch>:]PULSe:RECeiver:B<rec_no>:SRCPort<port_no>:TRIGger:DELay
<shift_stimulus>
Defines an offset time (shift stimulus) for the wave quantity assigned to the measurement receiver
<rec_no> and the analyzer source port <port_no>.
Number of the measurement receiver (= port number of the analyzer). Values
range from 1 to the total number of test ports.
Number of the analyzer source port. 1 to the total number of test ports.
–0.003 s to +0.003 s [s]. The increment (UP/DOWN) is 50 ns.
Device-specific, command or query
See [SENSe<Ch>:]PULSe:RECeiver:B<rec_no>:SRCPort<gen_no>:EVALuation:MODE
[SENSe<Ch>:]PULSe:TIME:BWIDth[:RESolution] <bandwidth>
Selects the IF bandwidth for pulse profile measurements.
100 kHz to 10 MHz [Hz]. UP and DOWN increment/decrement the bandwidth in 1-
2-5 steps for each decade (the value 3 MHz is provided in addition). The analyzer
rounds up any entered value between these steps and rounds down values
exceeding the maximum bandwidth.
Device-specific, command or query
*RST; SWE:TYPE PULS; :PULS:TIME:BWID 5 MHz