R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 958
Device-specific, command or query
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1.
SEGM:SWE:TIME 0.1
Increase the segment sweep time to 0.1 s.
SEGM:SWE:TIME:CONT OFF
Couple the sweep times in all segments and reset the sweep time in segment no. 1
to the initial value.
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:SUM?
Returns the total duration of the segmented sweep, calculated as the sum of the sweep times of the
individual segments ([SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME).
Sweep segment number. This suffix is ignored; the command returns the sum
of all segments.
See [SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME [s]
Device-specific, command or query
See [SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
[SENSe<Ch>:]SEGMent<Seg>:TRIGger:CONTrol <Boolean>
Enables/disables selective segment triggering (configured using
[SENSe<Ch>:]SEGMent<Seg>:TRIGger:STATe).
Sweep segment number.
This suffix is ignored: the command affects the whole segmented sweep.
OFF: Disables selective segment triggering.
ON: Enables selective segment triggering.
Device-specific, command or query