Instrument Functions
R&S
®
FSVA/FSV
381Operating Manual 1176.7510.02 ─ 09
This parameter is only available if the limit check is set to "Absolute" (see "Limit Check
1-4" on page 380).
Remote command:
[SENSe:]LIST:RANGe<range>:LIMit:STOP on page 820
Close Sweep List ← Sweep List ← Spurious Emissions
Closes the "Sweep List" dialog box and updates the measurement results.
Insert before Range ← Sweep List ← Spurious Emissions
Inserts a new range to the left of the currently focused range. The range numbers of
the currently focused range and all higher ranges are increased accordingly. The maxi-
mum number of ranges is 20.
For further details refer to "Ranges and Range Settings" on page 358.
Insert after Range ← Sweep List ← Spurious Emissions
Inserts a new range to the right of the currently focused range. The range numbers of
all higher ranges are increased accordingly. The maximum number of ranges is 20.
For further details refer to "Ranges and Range Settings" on page 358.
Delete Range ← Sweep List ← Spurious Emissions
Deletes the currently focused range, if possible. The range numbers are updated
accordingly. For further details refer to "Ranges and Range Settings" on page 358.
Remote command:
[SENSe:]LIST:RANGe<range>:DELete on page 816
Adjust Ref Lvl (span > 0) ← Sweep List ← Spurious Emissions
Adjusts the reference level to the measured total power of the signal. The softkey is
activated after the first sweep with active measurement of the occupied bandwidth has
been completed and the total power of the signal is thus known.
Adjusting the reference level ensures that the signal path will not be overloaded and
the dynamic range not limited by too low a reference level. Since the measurement
bandwidth for channel power measurements is significantly lower than the signal band-
width, the signal path may be overloaded although the trace is distinctly below the ref-
erence level. If the measured channel power is equal to the reference level, the signal
path cannot be overloaded.
Remote command:
[SENSe:]ADJust:LEVel on page 777
Adjust X-Axis ← Sweep List ← Spurious Emissions
Adjusts the frequency axis of measurement diagram automatically so that the start fre-
quency matches the start frequency of the first sweep range, and the stop frequency of
the last sweep range.
List Evaluation ← Spurious Emissions
Opens a submenu to edit the list evaluation settings.
For more information on list evaluation see "Result Evaluation" on page 374.