Instrument Functions
R&S
®
FSVA/FSV
380Operating Manual 1176.7510.02 ─ 09
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:GAIN:STATe on page 819
Sweep Points ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the number of sweep points for the specified range.
Remote command:
[SENSe:]LIST:RANGe<range>:POINts on page 820
Stop After Sweep ← Sweep List dialog box ← Sweep List ← Spurious Emissions
This command configures the sweep behavior.
"ON"
The R&S FSVA/FSV stops after one range is swept and continues
only if you confirm (a message box is displayed).
"OFF"
The R&S FSVA/FSV sweeps all ranges in one go.
Remote command:
[SENSe:]LIST:RANGe<range>:BREak on page 815
Transd. Factor ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets a transducer for the specified range. You can only choose a transducer that fulfills
the following conditions:
●
The transducer overlaps or equals the span of the range.
●
The x-axis is linear.
●
The unit is dB.
Remote command:
[SENSe:]LIST:RANGe<range>:TRANsducer on page 822
Limit Check 1-4 ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the type of limit check for all ranges. Possible states are:
Absolute Checks the absolute limits defined.
None No limit check performed.
The limit settings are only available if limit check is activated ("Absolute"). If activated,
the limit line and the results of the check are indicated in the diagram.
Remote command:
[SENSe:]LIST:RANGe<range>:LIMit:STATe on page 820
CALCulate<n>:LIMit<k>:FAIL? on page 638
Abs Limit Start ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets an absolute limit value at the start frequency of the range [dBm].
This parameter is only available if the limit check is set to "Absolute" (see "Limit Check
1-4" on page 380).
Remote command:
[SENSe:]LIST:RANGe<range>:LIMit:STARt on page 819
Abs Limit Stop ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets an absolute limit value at the stop frequency of the range [dBm].
Measurement Functions