Remote Control
R&S
®
FSVA/FSV
820Operating Manual 1176.7510.02 ─ 09
Example:
LIST:RANG1:LIM:ABS:STAR 10
Sets an absolute limit of 10 dBm at the start frequency of the
range.
Manual operation: See "Abs Limit Start" on page 380
[SENSe:]LIST:RANGe<range>:LIMit:STATe
This command activates or deactivates the limit check for all ranges.
Suffix:
<range>
.
irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
LIST:RANG3:LIM:STAT ON
Activates a limit check for all ranges.
Manual operation: See "Limit Check 1-4" on page 380
[SENSe:]LIST:RANGe<range>:LIMit:STOP <Limit>
This command sets an absolute limit value at the stop frequency of the specified
range. Different from manual operation, this setting is independent of the defined limit
check type.
Suffix:
<range>
.
1...20
range
Parameters:
<Limit> -400 to in 400 dBm
*RST: 13 dBm
Example:
LIST:RANG1:LIM:STOP 20
Sets an absolute limit of 20 dBm at the stop frequency of the
range.
Manual operation: See "Abs Limit Stop" on page 380
[SENSe:]LIST:RANGe<range>:POINts <Value>
This command sets the number of sweep points for the specified range.
Suffix:
<range>
.
1...20
range
Remote Control – Commands