Remote Control
R&S
®
FSVA/FSV
722Operating Manual 1176.7510.02 ─ 09
Note: The measurement is performed on the trace marker 1 is positioned. In order to
evaluate another trace, marker 1 must be positioned on another trace with
CALCulate<n>:MARKer<m>:TRACe.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:SUMM:PPE ON
Switches on the function.
Manual operation: See "Peak" on page 385
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:REFerence:AUTO ONCE
This command sets the currently measured average value (CALCulate<n>:
MARKer<m>:FUNCtion:SUMMary:MEAN[:STATe]) and RMS value
(CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe]) as reference
values for relative measurements in zero span.
If the measurement of RMS value and average is not activated, the reference value 0
dBm is used.
If the function CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage or
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd is switched on, the cur-
rent value is the accumulated measurement value at the time considered.
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Example:
CALC:MARK:FUNC:SUMM:REF:AUTO ONCE
Takes the currently measured power as reference value for the
relative power measurement in zero span.
Usage: Event
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:AVERage:RESult?
This command queries the average RMS time domain power. The query is only possi-
ble if averaging has been activated previously using CALCulate<n>:MARKer<m>:
FUNCtion:SUMMary:AVERage.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
Remote Control – Commands