List and sweep mode
R&S
®
SMB100B
119User Manual 1178.3711.02 ─ 08
Extern Start/Stop mode (sweep)
Figure 6-6: Extern Start/Stop mode (sweep)
The instrument generates the signal continuously.
Trigger mode "Auto" and "Sweep > State = On" are prerequisite.
State "On" sets the signal to the start value, that is one of the following:
Sweep start frequency
Sweep start power
Frequency-power value pair of the selected index in the list
Starts signal generation with a trigger event.
Switches automatically to the next sweep step when the Dwell time has elapsed.
If the end value is reached, signal generation continues with the next sweep cycle.
Stops signal generation with the next external trigger event.
Starts the signal generation again with the next trigger event, beginning at the start
value.
"State = Off" stops the signal generation in sweep or list mode.
Trigger source: An externally applied trigger signal.
Table 6-4: Cross-reference between manual and remote control in Extern Start/Stop modes (sweep)
Manual control mode:
"Extern Start/Stop"
Remote commands
Rohde & Schwarz proprietary and [SCPI conform] value name
RF frequency sweep
:TRIGger<hw>:FSWeep:SOURce EAUTo [-]
:SOURce<hw>:SWEep:FREQuency:MODE AUTO
RF level sweep
:TRIGger<hw>:PSWeep:SOURce EAUTo [-]
:SOURce<hw>:SWEep:POWer:MODE AUTO
LF frequency sweep
:TRIGger<hw>:LFFSweep:SOURce EAUTo [-]
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE AUTO