List and sweep mode
R&S
®
SMB100B
120User Manual 1178.3711.02 ─ 08
Manual mode (Sweep/List)
The manual mode only applies to remote control. It is not visible in the graphical user
interface of the instrument and is described here for completeness.
Figure 6-7: Manual mode in remote control (sweep / list )
●
The instrument generates the signal in steps.
●
Starts signal generation with "State = On".
●
Switches to the next step with a user defined setting via remote control.
You can arbitrarily select a value within the range of the start and stop values by
setting the frequency, power or index using the corresponding remote control com-
mand. There is no ascending or descending order.
●
Stops signal generation with "State = Off".
Table 6-5: Remote control commands in manual mode (Sweep/List)
Remote control mode:
Manual
Remote commands
Rohde & Schwarz proprietary and [SCPI conform] value name
RF frequency sweep
:SOURce<hw>:FREQuency:MANUal <Manual>
RF level sweep
:SOURce<hw>:POWer:MANual <Manual>
LF frequency sweep
:SOURce<hw>:LFOUTput:FREQuency:MANual <Manual>
List
:SOURce<hw>:LIST:INDex <Index>
Signal generation and triggering in the sweep and list modes