Improving level performance
R&S
®
SMB100B
182User Manual 1178.3711.02 ─ 08
a) Select "Select Column To Fill > Frequency / Hz".
b) Select "Range > e.g. 15" to determine the number of values.
c) Select "Start Value > e.g. 1.56 GHz".
d) Select "Increment Value > e.g. 1 MHz", to determine the frequency steps.
e) Select "Fill", to insert the frequency values.
The "Fill" function also fills the column of the "Correction Value / dB" values with a
predefined value, since empty cells lead to the data loss of the entire line. If you
need different values, you can change them manually, or you can use the auto-
matic function "Fill with Sensor" as described in the next steps.
8. Select "Fill with Sensor".
The "Fill User Correction Data With Sensor" dialog provides an overview of the
sensor configuration.
9. Select "Execute".
The R&S SMB100B successively sets each frequency point, reads the measured
power of the sensor and fills in the value in the correction table.
How to calibrate the power level with an R&S NRP power sensor