Improving level performance
R&S
®
SMB100B
183User Manual 1178.3711.02 ─ 08
10. Select "Save" to save the data in the file.
11. Close the dialog.
To perform power leveling calibration with user correction data
We assume that a user correction file is available in the user directory of the
R&S SMB100B or on a memory stick or in a shared directory.
If you have created and saved the file immediately before this step, the file is loaded in
the "User Correction" dialog automatically. Otherwise you can load a previously saved
file.
1. Select "Level" > "User Correction".
2. Select "UCOR Data" > "UCOR Data...", if there is no file loaded already.
3. Select the directory and file you want to use.
4. Load the file with "Select".
5. To view the file content, select "Edit UCOR Data...".
6. Select "UCOR Data" > "State" > "On" to apply the user correction values.
When you activate the RF output, the R&S SMB100B considers the user correction
data and adjusts the signal level accordingly to compensate external frequency
responses.
How to calibrate the power level with an R&S NRP power sensor