Remote control commands
R&S
®
SMB100B
478User Manual 1178.3711.02 ─ 08
[:SOURce<hw>]:LIST:DWELl:MODE <DwelMode>
Selects the dwell time mode.
Parameters:
<DwelMode> LIST | GLOBal
LIST
Uses the dwell time, specified in the data table for each value
pair individually.
GLOBal
Uses a constant dwell time, set with command [:
SOURce<hw>]:LIST:DWELl.
*RST: GLOBal
Example: See Example"Create an RF list and activate the list mode"
on page 474.
Manual operation: See "Dwell Time Mode" on page 140
[:SOURce<hw>]:LIST:DWELl:LIST
<Dwell>
Enters the dwell time values in the selected list in µs.
Parameters:
<Dwell> <Dwell#1>{, <Dwell#2>, ...} | block data
You can either enter the data as a list of numbers, or as binary
block data. The list of numbers can be of any length, with the list
entries separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy. See also :
FORMat[:DATA] on page 387 for more details.
Example: See Example"Create an RF list and activate the list mode"
on page 474.
Manual operation: See "Edit List Mode Data" on page 146
[:SOURce<hw>]:LIST:DWELl:LIST:POINts?
Queries the number (points) of dwell time entries in the selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example"Create an RF list and activate the list mode"
on page 474.
Usage: Query only