EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - Page 525

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
525User Manual 1178.3711.02 ─ 08
Return values:
<NextInfo> string
Example:
:SYSTem:NINFormation?
Queries the oldest entry in the info message queue.
Response: 90,"Info;=== Instrument startup...
==="
Information message containing error number 90, that states,
that the instrument startup is complete.
Usage: Query only
:SYSTem:ULOCk
<Mode>
Locks or unlocks the user interface of the instrument.
Parameters:
<Mode> ENABled | DONLy | DISabled | TOFF | VNConly
ENABled
Unlocks the display, the touchscreen and all controls for the
manual operation.
DONLy
Locks the touchscreen and controls for the manual operation of
the instrument. The display shows the current settings.
VNConly
Locks the touchscreen and controls for the manual operation,
and enables remote operation over VNC. The display shows the
current settings.
TOFF
Locks the touchscreen for the manual operation of the instru-
ment. The display shows the current settings.
DISabled
Locks the display, the touchscreen and all controls for the man-
ual operation.
*RST:
n.a. (factory preset: ENABled)
Example:
:SYST:ULOC DIS
Activates the user interface lock, including display and controls.
Manual operation: See "User Interface" on page 254
See "Enabling a locked user interface for manual operation"
on page 255
:SYSTem:LOCK:OWNer?
Queries the sessions that have locked the instrument currently.
If an exclusive lock is set, the query returns the owner of this exclusive lock, otherwise
it returns NONE.
SYSTem subsystem

Table of Contents

Other manuals for Rohde & Schwarz R&S SMB100B