Loading...
Remote control commands
R&S
®
SMB100B
526User Manual 1178.3711.02 ─ 08
Return values:
<Owner> string
Example:
SYST:LOCK:OWN?
Returns the owner of locking.
Response: NONE
The instrument is not locked.
Usage: Query only
:SYSTem:LOCK:RELease:ALL
Revokes the exclusive access to the instrument.
Usage:
Setting only
:SYSTem:LOCK:REQuest[:EXCLusive]?
Queries whether a lock for exclusive access to the instrument via ethernet exists. If
successful, the query returns a 1, otherwise 0.
Return values:
<Success> integer
Example:
SYST:LOCK:REQ?
Queries the state of exclusive locking.
Response: 1
The exclusive locking is active.
Usage: Query only
:SYSTem:SAV <Pathname>
Saves the current R&S SMB100B settings into a file with defined filename and into a
specified directory. The file extension (*.savrcltxt) is assigned automatically.
Setting parameters:
<Pathname> string
Example:
SYSTem:SAV "/var/user/temp/Test"
Saves the file Test.savrcltxt into the direc-
tory /var/user/temp/.
Usage: Setting only
:SYSTem:RCL <Pathname>
Loads a file with previously saved R&S SMB100B settings.
Loads the selected file with previously saved R&S SMB100B settings from the default
or the specified directory. Loaded are files with extension *.savrcltxt.
SYSTem subsystem