Web UI Page Current Status Indication Troubleshooting
HOME page Reference
Status Table
One or more
input ref-
erences indic-
ate “Not
Valid” (red)
All available Input References
have been lost. The Reference
Status table on this same page
will show the current status of
all inputs (Green is valid and
red is not valid, or not present).
If Signature Control is enabled
in this state, the output may be
disabled.
1. Make sure the Input Refer-
ence Priority table still has the
desired inputs enabled, based
on desired priority.
2. Make sure desired input ref-
erences are still connected to
the correct input port of
Ver-
saSync
.
3. Verify GNSS antenna install-
ation (if applicable).
Navigate to
INTERFACES/OUTPUTS/
PPS Output page
Select the PPS
Output
screen.
Signature Con-
trol will show
"Output
Always
Enabled",
"Output
Enabled in Hol-
dover",
"Output Dis-
abled in Hol-
dover" or
"Output
Always Dis-
abled".
1. With "Output Always
Enabled" selected, the selected
output will be present no mat-
ter the current synchronization
state.2. Any other configured
value will cause the applicable
output to be halted if
Ver-
saSync
is not fully synchronized
with its input references.
1. The PC(s) must be routable to VersaSync. Make sure you can access Ver-